| Clinical question | Does my patient have elevated ICP? |
|---|---|
| Patient positioning | Supine |
| Supplies needed | Small tegaderm (to avoid ultrasound gel irritating the cornea) |
| Probe | Linear |
| Normal findings | The upper limit of normal optic nerve sheath diameter: 4 mm in infants. |
| 5 mm in children. | |
| 5 mm in adults. | |
| Sensitivity & specificity | A meta-analysis of 12 studies showed a sensitivity of 95.6% and specificity of 92.3% |
| Contraindications | Open globe injury, periorbital swelling or hematoma, ocular trauma, history of glaucoma |
| Alternate diagnostic modalities | Fundoscopy, non-contrast CT scan |


Note that the optic nerve diameter is being measured ~3mm from the retina.

Axial & Sagittal (incorrectly labeled coronal above) views of the optic nerve sheath shown here